Benchmark dataset for validating computational fluid dynamic (CFD) simulation of blood flow through FDA Nozzle and FDA blood pump /

Computational benchmark #1 (Nozzle)

by Prasanna Hariharan

PROJECT HOME FDALink

nozzle.png

For the first study, a bidirectional nozzle model was created in which the flow diameter changed either suddenly or gradually. Particle image velocimetry measurements were acquired in three laboratories using a Newtonian blood analog fluid which matched the refractive index of the nozzle models. The flow rate was adjusted so that the velocity field and pressure along the model could be measured at Reynolds numbers in the nozzle throat which corresponded to laminar (Re = 500), transitional (Re = 2000), and turbulent flow regimes (Re = 3500, 5000, 6500).

Details about geometry, flow conditions, experimental dataset, publications from the working group, acknowledgements, and citations using the benchmarks are provided in the links below